Extractables and Leachables from Single Use Systems and Multi-Use Equipment Used in Pharmaceutical and Biotherapeutic Manufacturing

I.         Abstract

Leachables from single use systems (SUS)and multi-use manufacturing equipment (MUE) that migrate into drug or biological products have a potentially negative impact on safety. To evaluate this risk, extraction studies are performed under conditions designed to simulate the use of the equipment.   The goal of these extraction studies is to identify as extractables the leachables that could migrate into the product.  The extractables profile can then be used to set operational limits on the SUS or MUE as needed.   Further toxicological assessment may be required based upon the observed levels of leachables.

II.       Introduction

Leachables are compounds that migrate into drug or biological products from materials encountered in the manufacturing process.   Plastics and elastomers components are the most common sources of organic leachables and metal components are the most common sources of inorganic leachables.  Examples of common organic leachables can be seen in Table 1.  Any component of the manufacturing equipment with direct contact with a liquid that eventually becomes part of the final product is a potential source of leachables.  Examples of SUS and MUE components that can give rise to extractables include filters, hoses, bioreactor bags, connectors and filling needles.   Manufacturing equipment that is not directly in contact with a liquid that enters the final product is unlikely to be a source of leachables.    Leachables present a potential risk to the patient both from the toxicity of the leachable and from the possible negative impact upon stability and efficacy of the final product.  

The manufacturers of drugs, biologics and biotherapeutics are required to evaluate the risk of leachables from SUS and MUE used in the manufacturing process. Many equipment vendors will provide extractables results for their components generated in many different model solvent systems that bracket the pH, polarity, and ionic strength of most aqueous based drug products.  The vendor extractables testing is typically performed according to the “white papers” issued by the Bioprocess Systems Alliance (BPSA) and the BioPhorum Operations Group (BPOG). A scientific based assessment comparing the extractive power of the drug product that will be manufactured using the equipment to the vendors conditions of contact in the extractables testing has to be performed to demonstrate that the vendor extractables profiles were generated under “worst case” conditions.  Any “gaps” identified in the correlation of the vendor extractables data to the conditions used in manufacturing can bead dressed utilizing a simulated-use extraction study.  As part of the assessment, a justification can be made to exclude components of the manufacturing train from the simulated-use extraction study if there will be significant dilution of leachables or potential removal of the leachables during purification techniques (such asrecrystallization and diafiltration).

In a simulated-use extraction study, the components of the SUS or MUE are extracted with a model solvent or placebo under representative conditions of contact encountered during manufacturing.  The sample extracts are analyzed by GC-MS, LC-MS and ICP-MS to identify all of the organic and inorganic. Any pre-treatment of the components that would be done during manufacturing (e.g. pro-active flushing of filters and tubing) should be done on the sample before the start of the simulated-use extraction study.

 

I.       Simulated Use Extraction

A.    Sample Extraction

In a simulated use extraction study, the extraction solvent is selected to closely mimic the drug product.   The extraction conditions are usually static storage at the temperature intended for temperature for a minimum of two time points.   The first time point is the intended routine time the individual component would be in contact with the product and the final time point would represent the longest allowable time the product would be in contact with the product.

B.    Extractable Analysis

The sample extracts are analyzed by GC-MS,LC-MS and ICP-MS.   The goal of these analysis is to identify as many extractables as possible and to semi-quantitatively determine the level of each extractable.  Since the methods are designed to detect unknowns, these methods cannot be validated. Results from these analyses are reported as the amount of the extractable (usually in µg) per weight (usually in g) or surface area (usually in cm2) of the component.

C.  Analytical Evaluation Threshold

At the completion of the extractables analysis, a list of extractables is generated.  The challenge at this point is to only report extractables that present a toxicological risk.   To evaluate the toxicity of each observed extractable, the safety concern threshold (SCT) is used.  

 

The SCT is the absolute highest acceptable exposure of a patient to a leachable in drug product and is usually expressed in terms of µg of leachable per day.  If an SCT is not known, the recommended SCT by the Product Quality Research Institute (PQRI) for a parenteral drug is 1.5 µg of each individual leachable per day (1).   The PQRI selected this SCT as representing a threshold below which leachables would have negligible safety concerns from carcinogenic and non-carcinogenic toxic effects.  

 

To apply the SCT to a given product, an analytical evaluation threshold (AET) is calculated based on the SCT of an individual leachable (usually in µg) per day, the number of doses of the drug product administered per day, the number of doses contained in the container closure system, and the volume of product in the container closure system .   The AET is defined as follows:

The AET will have units of µg/mL of the product and can be converted to the sample extracts based upon the extraction conditions.

 

D.    Interpretation of Results

If quality manufacturing equipment has been selected and the extraction conditions were appropriately selected, the number of extractables observed will usually be small.   From the generated extractables profile,only the extractables above the AET present a risk to patient safety.  

As is commonly observed, in some cases the levels of extractables will increase over time until it rises above the AET.   If a time can be identified where the extractables levels are below the AET, the manufacturing process should be modified to limit the exposure time thus ensuring the extractable levels will stay below the AET in the final product.   This is commonly sufficient to eliminate the risk from leachables from manufacturing components.

In rare cases, changes to the manufacturing process are not sufficient to reduce the risk from a specific extractable.    In this case a toxicological assessment to determine a compound specific SC is recommended to fully evaluate the risk from the specific extractable.

 

I.       Conclusion

 

Leachables present a unique challenge in assuring drug product safety and efficacy.   The experimental approach presented represents a rational experimental approach to evaluate this risk.  When properly executed and interpreted, a simulated use extraction can be used to evaluate and minimize the risk of leachable from manufacturing equipment.  

 

II.     Acknowledgements

The authors would like to thank Michael Ruberto of Material Needs Consulting for his contributions to this white paper.

 

VI.      References

 

1.    “TheProduct Quality Research Institute (PQRI) Leachables and Extractables Working Group Initiatives for Parenteral and Ophthalmic Drug Product (PODP)”, PDA JPharm Sci and Tech 2013, 67 430-447

READ
October 2, 2019

Extraction Study of a Cartridge Filter Based Upon the BioPhorum Operations Group (BPOG) Recommendation

READ
May 31, 2019

Simulated Use Extraction as an Alternative to Leachables Testing

If good materials were selected for the container closure system (CCS) for a parenteral drug product, it is not uncommon for the results of the forced extraction study to be that no extractables were detected above the analytical evaluation threshold (AET).    While the PRQI guidances does allow for the conclusion to drawn from these results that the CCS presents no risk to patient safety, many drug development teams are hesitant to make this conclusion due to the legitimate concern that regulatory reviewers may still request leachables results.    However,  the same drug development team may be hesitant to commit the time and money for unneeded leachables testing.    An alternative study proposed by Pine Lake Laboratories is a simulated use extraction to further support the conclusion of no risk from leachables without the expense and time of full leachables testing.

In a simulated use extraction study, the extraction solvent is selected to closely mimic the pH and solvating strength of the drug product.    For simple drug products, the placebo may be a good choice if it does not present significant interferences to the analytical methods used to detect the leachables.   The extraction solvent should be selected to ensure it is compatible with the analytical methods.

The extraction conditions are selected based upon the intended shelf life of the drug product so that the “real time” storage equivalent can be determined.    According to USP <1663>, the relationship between the diffusion rate and temperature can be expressed empirically by the Arrhenius equation and has been established in ASTM F1980-07 (2011) “Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices”.     Therefore, the following equation can be used to determine the “real time” storage equivalent for a drug product intended to be stored at  5 C from the results of a simulated use extraction where the CCS is extracted for 60 days at 40 C :


AAR (Accelerated Aging Rate) = Q10 ((Te – Ta)/10)


Where Ta = Normal Storage Temperature = 5 C

Te = Elevated Temperature = 40 C

Q10 = Reaction Rate = 2 

= 2 ((40 –5)/10)

= 11.31

AATD (Accelerated Aging Time Duration) x AAR = Desired Real Time 

60 Days x 11.31 = 679 days or 22.6 Months at 5oC


As can be seen, a simulated use study done at 40 C for 60 days for a drug product intended to be stored at 5 C would give the same results 22.6 months at the target storage condition.   The time and temperature of the simulated use extraction can be altered as need based upon the targeted shelf life and storage conditions.

At the end of the simulated use extraction study, the extractables profile generated will be for an extraction solvent that closely mimicked the drug product under extraction conditions that are equivalent to the intended shelf life and storage conditions.    This extractables profile provides valuable additional information that can be used to ensure patient safety without the need for full leachables testing.

READ
January 4, 2019

Why Choose a Small CRO over a Top 10 CRO?

‘Our Mission is to work with clients to assist in bringing innovative, safe new medicines and medical devices to patients in need.’

Pine Lake Laboratories is a small GLP/GMP compliant analytical and bioanalytical contract testing laboratory that is FDA registered located in Bristol, CT. Pine Lake Laboratories has extensive experience with method development and validation and specializes in early development work. Pine Lake Laboratories has the capabilities to help pharmaceutical, medical device, and contract manufacturing companies get their product to market faster and more efficiently.

Contract Research Organizations that promote “one stop shop” services with various locations are competitors and are sometimes chosen as the ‘safe option’ due purely on total number of resources. However, choosing a large scale CRO comes with problems of its own; stakeholder ambiguity, increased regulatory scrutiny, and ridged bureaucracy.

Small or midsize projects become less of a priority to companies with the most purchasing power. We give unwavering service across the board, no matter your size or name.

Choosing a smaller contract research organization comes with advantages; flexibility, swift decision making, personalization and communication. There is no barrier between client and President of the company, or large steps from sales to scientist. All employees have a voice and can help the client succeed in getting their product to market safely.

Flexibility

Pine Lake Laboratories understands that the industry is continuously growing and innovating. We strive to be as flexible and adaptive for our clients because we know they are fully immersed. For example, a client may be informed by their vendor of a upcoming material change in the bio-processing system and we will pivot our study plan accordingly for this new material.

Moreover, being a small business, we accept projects that can be seen as a “risk” to large companies; a project that takes a lot of resources due to it being more ‘experimental’ in nature. Pine Lake Laboratories will listen to your companies needs and use our expertise to come up with workable solutions to meet your budget and your teams’ goals.

Pine Lake Laboratories is a dedicated partner in all of the projects we work on and we will remain fully engaged until our customers are satisfied.

Innovation/ Accommodating

The pharmaceutical industry is constantly changing, from new ideas, regulatory guidelines, and best practices. Large CROs could have a hard time adjusting to change, small CROs don’t focus on the high throughput, rinse and repeat strategies. Pine Lake Laboratories will go the extra mile for your product, they have the instrumentation and knowledge to complete your testing and create a data package that can be submitted directly to the FDA.

All customer requests are explored and only accepted if we bring value to a project. Each client is given a personalized study design that will satisfy all their goals for their product. If a service is needed that Pine Lake Laboratories cannot fulfill, we can either outsource or make a referral to an appropriated trusted partner so that our client has piece of mind.

Furthermore, we are able to adapt new study designs quickly and smoothly with our well versed cross-trained experts and all without the unproductive layers of red tape.

Employee Input

Employees from Pine Lake Laboratories come from diverse scientific backgrounds. They bring experience from both small and large pharmaceutical and biotech companies. Pine Lake Laboratories is confident in our abilities to bring the correct technical expert to the discussion table to provide our clients with a full range of GMP and GLP analytical solutions.

All of our employees are dedicated to establishing trust, integrity and delivering results required to build lasting client relationships. Clients can be assured they will not just be another number with Pine Lake Laboratories.

Moreover, Pine Lake Laboratories specializes in services such as CMC Analytical, Extractables and Leachables, Bioanalysis and more. Our equipment required for our service offerings is state of the art. Moreover, our employees are not siloed into just one technique.

All employees participate in the continuous learning program at Pine Lake Laboratories, in order to refresh or educate all employees on leading industry practices. Not one scientist is a specialist on one service or instrument; they are all experts, highly trained on all services and instrumentation.

Personalization

At Pine Lake Laboratories our clients are treated as coworkers. We take time to understand their goals and mission. Their mission statement becomes a part of ours; clients can express their visions, struggles, and outlooks. A call or visit from the President of Pine Lake is only a simple request away and easily accommodated. Dr. Kurt Moyer is happy to visit clients’ offices to meet project teams and fully explain the capabilities and services Pine Lake Laboratories has to offer. Furthermore, visits to Pine Lake Laboratories are also welcome- before or during projects. Clients may want to take a tour of the facilities, meet the team, conduct an audit, or visit during a specific milestone of their testing.

Pine Lake Laboratories makes every step of the process as painless as possible; from quote generation to final report. Again, they can easily speak with our key employees to get any answers to their questions.

Conclusion

Pine Lake Laboratories is a GLP/GMP compliant analytical and bioanalytical contract testing organization that has the capabilities to solve even the most challenging of analytical problems by partnering with our clients, being adaptive, innovative and accommodating, as each client is a valuable member of our team.

READ
November 3, 2018

Extractables and Leachables Testing

Pine Lake Laboratories is the scientific leader in extractables and leachables (E&L) studies.   We have extensive experience with pharmaceutical container closures, single use system components, and combination devices.

Our experienced scientific staff utilizes the following analytical methods for expert identification of extractables observed in a forced extraction study:

  • Headspace GC-MS for identification of volatile organic extractables
  • GC-MS for identification of semi-volatile organic extractables
  • LC-QToF for identification of non-volatile organic extractables
  • ICP-MS for identification of inorganic extractables
  • Polynuclear Aromatic Hydrocarbon Screening (GC-MS)
  • Nitrosamine analysis by GC-NCD

Our team has experience developing and validating leachables methods based upon the observed extractables in a wide variety of dosage forms:

  • Orally inhaled and nasal drug products (OINDPs)
  • Parenterals (small and large volume)
  • Prefilled syringes
  • Opthamalic
  • Topicals
  • Transdermals
  • Implantable medical devices
  • Combination medical devices

Protocols are designed to be study specific and follow PQRI, BPOG or client specific.  Pine Lake Laboratories can also perform USP <661.1> and <661.2>.

A toxicological risk evaluation for E&L results can be performed by our staff toxicologists.

READ
November 2, 2018

Quality Agreements

INTRODUCTION

In the present article we summarize what we’ve learned in the period since our earlier article appeared in Journal of GXP Compliance some years ago, entitled “Implementing Quality Agreements at the Contract Laboratory”1. Specifically, here we focus on how the lab gets the Quality Agreement “right” and then follows it throughout its lifecycle by identifying what we call key “check points”. Our experience has shown that there are points in the Quality Agreement process, by which we mean the process of drafting, approving, implementing and revising the Quality Agreement, that are particularly problematic because they require greater attention and/or pose greater risk and it is key to manage these.

Check Point ONE: Prepare to adopt agreements

A major challenge at the contract lab (CRO) is to handle Quality Agreements (QAGs) through some standardized procedure, while at the same time allowing for the diverse needs of a wide client base.  This becomes more evident when one considers that the QAG typically originates with the client, in their format, and therefore contains many client-specific idiosyncrasies. Furthermore, the contract lab may be performing laboratory testing under GLP, GMP and/or ISO laboratory testing protocols, which, although greatly similar to one another, will each have its own unique requirements. Procedures, to be useful, must be detailed enough for personnel to be trained on and to follow. However, to be standardized across the wide client base, they must not be too detailed.  One cannot have a procedure for each client. Management must therefore “see the forest for the trees”.

Check Point TWO: Check that your quality house is in order

There should already be in place a strong quality program at the CRO and the QAG must make a good fit to this program. Those concerns (terms) that are rather common to Quality Agreements should already be addressed in the existing quality program at the CRO or else consideration must be given to incorporating them into the existing program. This approach is distinctly different and better than leaving them only in the QAG. To leave important concerns only in the QAG is tantamount to instructing the lab to only consider doing such quality work when working on this particular client’s project.

Check Point THREE: Understand the high-risk terms of the specific agreement

Understand and have input into specific terms of the agreement. It is important to make a careful review of the draft. Don’t agree to terms that you cannot meet. Take a team approach in order to foresee any/all difficulties in implementing the terms by involving lab Management, Principal Investigators and Quality Assurance in the draft review. Many of the high-risk terms are those that are specific to the client. By high-risk, we mean at risk of being ignored during the conduct of the study.  For example, how and when the client is to be notified of an out-of-specification result is typically client-specific and terms involving notifications to be made to the client are generally high risk. Other terms that are often high risk center around how changes to laboratory procedures, testing protocols and written instructions in general are to be handled and how vendors and other 3rd parties providing materials and/or services in support of the client’s experimental study are to be qualified2.

Check Point FOUR: Train users on Quality Agreements in general and on each specific agreement they will be using

Quality Agreements are to be used. They are not to be put on the shelf and then forgotten. Keep track of who has been trained on each.  This can be done with a training matrix.  It is a good idea to keep an updated list of the agreements that are active (e.g. the training matrix) in a place highly-visible to all employees. To make it more useful, consider cross-referencing each QAG against the client and perhaps also to the Principal Investigators assigned to the client’s various experimental studies. To protect client-confidentiality, consider using a client code in place of the client name. A mechanism must be in place to ensure that the training is performed on a timely basis. Couple this training to some other required client-based activity that occurs at an appropriate time in the process of taking a client project from start to finish. For example, we start each experimental study using a kick-off meeting facilitated by the Principal Investigator (PI), who is guided by a kick-off meeting checklist3. Listed on the checklist is the requirement to review any applicable QAG. All scientists participating in the study are required to attend and to sign the attendance sheet. The attendance sheet is forwarded to the Training Coordinator, who uses it to update the Training Matrix. Any person missing the training must make it up and this will involve among other things, reading the QAG and passing a quiz on it. Attendance sheets are filed in the employees training binder.

Create an SOP on Quality Agreements. In it, provide background and details on the QAG system in place in general. All employees supporting client studies must be trained on this SOP.

Quality Assurance should keep the original QAG. QAGs are often 20 to 30 pages in length. It is a good idea for the QAU to also keep a one-to-two-page summary of key points appearing in the QAG and to note those points that the QAU anticipates causing more trouble to the lab or that may present a greater risk. The effective start and expiration dates along with client contact information should be included. The summary does not take the place of the QAG.

Check Point FIVE: Identify failures over time and feedback into the CAPA program

Any deviation from a QAG should be considered “major”, requiring a deviation investigation in order to identify root cause and corrective and preventative actions for Management and Quality Assurance review and approval4. The specific QAG will state whether or not the deviation is to be reported to the client and if so, how it is to be addressed with the client. Quality Assurance must query the CAPA database to determine if there is a pattern of failures to follow a specific QAG or QAGs, in general. Such feedback is critical to improving the QAG system and may help to improve specific Quality Agreements. Of course, the more effort put in at the front end to establish a good agreement and to train everyone on it, the less effort needed on the back end.  Perhaps more importantly, like any quality system, a good QAG system will scale up with no difficulty as more and more clients desire to put QAGs into place with the lab, whereas a bad system simply will not.

SUMMARY

The effective implementation of a single Quality Agreement and the management of a quality agreement program at the Contract Research Laboratory, requires getting good QAGs into place and monitoring the lab’s performance in implementing them over time. Key to the management process is the identification and understanding of points in the workflow at which to make critical checks.

REFERENCES

  1. 1. R.C. Wedlich and J. Webb, “Implementing Quality Agreements at the Contract

Laboratory”, Journal of GXP Compliance, Vol. 16, Issue 4, 2012.

  1. 2. R.C. Wedlich, et al., “Implementing an Effective Provider Qualification Program at the

Contract Laboratory”, Journal of GXP Compliance, Vol. 17, Issue 3, 2013.

  1. 3. R.C. Wedlich, et al., “Implementing an Effective GLP Program at the Contract Research

Laboratory”, Journal of GXP Compliance, Vol. 19, Issue 2, 2015.

  1. 4. R.C. Wedlich, et , “Implementing an Effective CAPA Program at the Contract

Laboratory”, Journal of GXP Compliance, Vol. 21, Issue 2, Mar 2017.

READ
October 19, 2018

Analytical Methods For Single Entity Combination Products

Introduction

The demand for sophisticated drug delivery has led to the development of many innovative combinations of pharmaceuticals, biologics and medical devices across a broad range of therapeutic areas. While these combination products have the potential to fulfill major unmet medical needs, they also present unique challenges in development and approval. One of these challenges is the development and validation of analytical methods to assess the quality, safety and stability of combination products.

Types of Combination Products

The FDA Guidance for Industry Titled, “Current Good Manufacturing Practice Requirements for Combination Products (January 2017)” defines a combination product as “a product composed of two or more types of medical products (i.e. a combination of a drug, device, and/or biological product with another).” There are three types of combination products: single entity, co-packaged and cross labeled.

A single entity combination product has two or more related components (i.e. drug/device, biologic/device, drug/biologic or all three) that are physically, chemically, or otherwise combined to produce a single entity. Examples of single entity combination products would be a prefilled syringes or a drug eluting stent.

A co-packaged combination product has two or more components packaged together in a single package. Examples of a co-packaged combination product are surgical and first aid kits.

With a cross labeled combination product, each component is packaged separately and is intended only for use with an approved second product. An example of a cross labeled combination product would be a light activated drug requiring a light emitting device.

Single entity combination products (SECP) present unique analytical challenges compared to co-package and cross labeled combination devices. Most notably, a SECP runs the risk of impurities and degradants from the combining process and sterilization. In addition, an SECP may have specific design features that need to measure to ensure proper functioning of the product.

As a result of these unique analytical challenges for SECPs, this white paper will focus only on the analytical methods needed to support the development of an SECP. For brevity, this white paper will only discuss SECPs formed from a drug and a medical device. Similar approaches with the appropriate methodologies could be adapted for the other types of SECPs.

Analytical Methods for Assay, Related Substances and Degradation Products

There are three goals for these methods: to assay the drug content, to measure for known related substances and to detect degradation products of the drug. Like the analytical methods for more traditional drug products, the predominant instrumentation for these methods is high-performance liquid chromatography (HPLC)- UV, with one method usually able to accomplish all three tasks.

The main difference between analytical methods for SECPs and traditional drug product formulations like oral or parenteral dosage forms is in the sample preparation. The sample preparation is unique and specific to the SECP but usually involves complicated multi-step procedures. Based upon the SECP, sample preparation of SECPs often involves extensive extraction for complete recovery of the drugs. In addition, depending upon the size of the SECP and the location of the drug, the SECP may need to be disassembled or reproducibly cut, which can be a significant challenge when the SECP is made from hard plastics or metals.

Since the method is intended to be used as to evaluate stability of the SECP, the method will need to be proven to be stability indicating. If the SECP contains a new drug, a forced degradation study will be needed. If the SECP contains a generic drug, literature references to known degradation pathways and impurities can be used to evaluate if the method is stability indicating. However, if the combining process is significantly different from the literature references, additional degradation pathways may need to be explored experimentally.

When designing a forced degradation study for a SECP, the potential for the device to contribute to the drug degradation or the degrade itself, leading to detectable degradation products needs to be considered. The type of SECP will determine if the forced degradation study is done on just the drug, on the fully assembled SECP or on a combination of the drug and the parts of the device with direct drug contact. It is recommended to include the components of the SECP that have direct drug contact in the forced degradation study.

The conditions of the forced degradation study may also need to be adapted for an SECP. Stress conditions can be replaced with more relevant experimental conditions with proper scientific justification.

Once developed, the analytical methods need to be validated according to the ICH Harmonized Tripartite Guideline , Validation of Analytical Procedures: Text and Methodology (ICH Q2(R1)). One important additional consideration for validation of methods for a SECP is during ruggedness. In the ruggedness validation experiments, additional challenges should be added to evaluate ruggedness of the sample preparation. These experiments will vary depending upon the SECP and need to appropriately challenge each critical sample preparation step.

Residual Solvents Methods for a SECP

If organic solvents are used in the assembly of the SECP, the finished SECP will need to be tested for these residual solvents. A common example would be if the drug is sprayed onto part of a medical device during the assembly of the SECP. In this case, the solvents used in the spraying process would be considered residual solvents.

The residual solvents methods need to be developed and validated in accordance with USP <467>. One important additional consideration for residual solvents for a SECP may be the need for additional sensitivity. Because these residual solvents may be important parameters in developing the process to manufacture the SECP, the methods may be needed to reach sensitivities as much as 10 fold lower than the required specification to assist in development of the manufacturing process. Once the methods have been validated, residual solvents testing can be done as part of release testing or as part of the process validation.

Analytical Method to Evaluate the Drug Release/Elution from a SECP

If the SECP is not intended to mechanically infuse the drug into the patient but instead the drug is intended to passively diffuse into the patient, then analytical methods designed to measure the rate of drug release or elution are needed. To perform this experiment, the SECP is placed in an appropriate media designed to model the target tissue. The analytical method is then used to measure the increase in drug concentration in the media over time. The instrument conditions used are usually similar to those used for the assay; however, more sensitive methods may need to be developed if the release of drug is expected to be slow. The method will need to be validated similar to an assay method with special consideration to sensitivity and additional ruggedness testing for the sample preparation factors that might impact the release date.

Analytical Method to Evaluate Uniformity

For SECPs where the medical device is coated with a drug or drugs, the uniformity of this coating will need to be evaluated. Similar to the drug release methods, the instrument conditions used are usually those used for the assay; however, more sensitive methods may need to be developed depending upon the intended level of drug in the coating. In addition, the sample preparation will need to be adjusted so that samples are taken from all areas of the device to ensure uniformity of the coating. In some SECPs where the device is not a simple geometric shape, the surface area coated from the different sections of the device will need to be included in the determination of the level of the drug in the coating.

Analytical Methods for the Analysis of Leachables from the SECP

Leachables (a.k.a. migrants) from the medical device need to be considered when the drug is in direct contact with the medical device during the intended shelf storage, when the SECP is intended to be surgically implanted or when the SECP will have direct patient contact for an extended period of time.

For these types of SECPs, two types of studies are performed. The first study is a forced extraction study on just the medical device and the second study a migration study on the entire SECP.

In a forced extraction study, the medical device is extracted with two solvents at an elevated temperature. Usually the drug is not included in the forced extraction. The extraction solvents are selected so that one mimics either the drug formulation or the intended patient tissue that the SECP will contact and the second solvent is selected to represent a “worse case scenario” condition based upon either the drug formulation or the intended patient tissue. The sample extracts are analyzed by mass spectrometry (gas chromatography, liquid chromatography, or inductively coupled plasma (GC-MS, LC-MS, and ICP-MS respectively)) to attempt to identify all possible organic and inorganic extractables.

Analytical methods are then developed that can detect the extractables observed in the forced extraction studies as leachables in either the drug product or a model solvent that mimics the intended patient tissue. Analytical methods for leachables need to be extremely sensitive and usually require MS detection. An additional challenge commonly arises when the drug product is present at concentrations significantly higher than the levels required for detection of the leachables. In this case sample preparation steps and method adaption need to minimize the interferences from the drug. Once developed, the analytical methods need to be validated before proceeding to the migration study. The validation of the methods should be similar to validation of a method for related substances but allowances may be needed to reach the required level of sensitivity.

The migration study is the second study where the leachables (a.k.a migrants) are monitored. When the risk of leachables is deemed to be highest from the drug being in direct contact with the medical device during the intended shelf storage, the leachables should be evaluated as part of the stability study to determine the shelf life. When the risk of leachables is due to the SECP being surgically implanted or from direct patient contact, a simulated migration study is performed. In this case the SECP is exposed to a model solvent that mimics the intended patient tissue at 37°C for an appropriate length of time determine by the intended use. In both studies, previously validated analytical methods are used to evaluate the leachables entering the drug product or the model solvent.

Other Analytical Methods

Since many types of SECPs are in development with many unique critical features, analytical methods may be required to measure qualities specific to a given SECP that are expected to be critical to function properly. In this case the analytical method will need to be developed to address a specific attribute of the SECP.

These methods will still require validation even when few of the standard validation parameters apply. In this scenario, the validation should address at a minimum reproducibility and ruggedness. Ruggedness testing should include all method parameters that could impact the reported results.

Conclusion

As diverse and ingenious SECPs continue to be developed, analytical methods that can be used to support the development and ensure the quality of these products are needed. A thorough understanding of the SECP is critical to ensure that the analytical methods are monitoring the proper attributes of the SECP and the analytical chemist may need to be creative in developing methods and preparing samples to support these important therapeutic advances.

READ
October 5, 2018

Short Chain Fatty Acid Analysis

To support microbiome research, Pine Lake Laboratories has developed an assay for short chain fatty acids (acetic, butyric, and propionic acid) in feces. These targeted metabolites can be used as an indicator of microbiome activity in research subjects and as biomarkers to evaluate the effect of various pharmaceutical drug treatments during pre-clinical and clinical trials. The method involves extracting the short chain fatty acids form feces then analysis by direct injection GC-MS.

This white paper will describe the method and highlight the performance of the method.

  1. Method Description

A. Sample Preparation

B. Standard and QA Preparation

Standards and QCs are prepared by spiking known levels of each short chain fatty acid into feces.   The endogenous level of each short chain acid in the feces used to prepare the standards and QCs had previously been determined by standard addition.   The final concentration is then the total of the amount added and the endogenous level.

Standards and QCs are then prepared for analysis same as the samples.

C. Instrumental Analysis

Samples are analyzed by GC-MS using a Nukol capillary column (15 m x 0.23 mm, 0.25 µm). Details available upon request.

  1. Results

A. Linearity and Range

For all three short chain fatty acids, the range of the method was 5 to 150 μg/mL with R2 > 0.990. Representative calibration curves can be found in Figures 1-3.

B. Precision in Matrix

A representative lot of feces was assayed in triplicate for all three short chain fatty acids.  Results are in Table 1. Precision on all matrix samples (n=6) was excellent with %RSD values less than 5% for each SCFA within each of the three analyses. The total precision of the grand average was less than 10% for acetic and propionic acid. Butyric acid was higher at 26.6% but was still acceptable at less than 30%.

C. Accuracy and Precision

A representative lot of feces was spiked at three separate levels with known amounts of each of the three short chain fatty acids.  This was repeated in three separate analyses. Results are in Table 2-4. Accuracy was acceptable for all three short chain fatty acids with average recovery within ± 15.0% of target for each level in each analysis and in the grand average. Similarly precision was acceptable at ≤15.0% for each level in each analysis and in the grand average.

  1. Conclusion

This simple yet effective GC/MS analysis can be used to accurately determine the concentration of acetic, butyric, and propionic acid in feces. The method is specific, accurate, and precise with a large quantifiable range for short chain fatty acids.    Additional short chain fatty acids not included in the method evaluation reported in this white paper can also be detected and quantitated using this method.

READ
September 20, 2018

Pharmaceutical Extractable and Leachable Studies

Abstract

Leachables from sample container closure systems (CCS) from primary and secondary packaging component that migrate into a drug products have a potentially negative impact on safety.  Analytical methods are needed to detect leachable in the drug product. The first step toward developing analytical methods for leachables is to identify the extractables that could become leachables by doing extraction studies. Extraction studies are designed to simulate both intended use and “worst case scenario” models to identify as extractables the leachables that could migrate into the drug product.  Analytical methods are then developed with sensitivity to detect the leachables in the drug product at the threshold determined by the toxicity of the leachable.  Analytical methods for leachables are validated similarly as methods intended to evaluate the stability of the drug product.  Leachable analysis can then be used in the long term stability studies or in migration studies designed to specifically evaluate leachables.

Introduction

Leachables are compounds that migrate into a drug product from the sample container closure system (CCS) under normal storage condition.  Both the primary CCS in direct contact with the drug product (metered dose inhaler, prefilled syringe, eye dropper, IV bag, HDPE bottle, LDPE ampoule, etc.) and the secondary CCS which does not contact the drug product (printed label, cardboard box, foil pouch, environmental exposure, etc.) can be sources of leachables. Leachables present a potential risk to the patient both from the toxicity of the leachable and from the possible negative impact upon stability and efficacy of the drug product.  Examples of common leachables can be seen in Table 1.

Although many types of materials can be used in a primary CCS system, the three most common are glass, polymers and elastomers.  One may expect that the manufacturer of the component of the CCS to be able to provide a complete list of the formulation and process used to manufacture the component, however this may not always be the case. The two main reasons for manufacturers not providing this information are that the manufacturer may consider the information to be proprietary or the manufacturer may not have the information. The second reason is particularly common for polymers. The main reason for this among the manufactures of polymer CCSs is that their upstream suppliers do not need to place strict control over their processes.  For example, a resin manufacturer will set specification for their product on its physical characteristics only and then sell the same resin to a manufacturer of a CCS and a manufacturer of lawn furniture. This example resin manufacturer may not have needed to keep accurate records on the amounts and type of antioxidants used as long as the resin met the manufacturer’s specifications, but these antioxidants do have the potential to leach into a drug product.

Leachables can enter any type of drug product including solid dosage forms. Generally, orally inhaled and nasal drug products (OINDP) and parenteral and ophthalmic drug products (PODP) are the most common drug products at high risk of leachables. Table 2 summarizes the risk for most common drug products. Low risk is not the same as no risk as evident in several high profile of recalls of solid dosage forms due to leachables. An assessment of the risk of leachables into a given drug product needs to be done when considering a testing strategy for leachables.

The toxicity of a leachable is dependent upon the route of entry into the body.  Levels of a compound that can be safely ingested can have a toxic effect when the same level is inhaled. As a result the potential route of administration of a leachable must be considered when assessing the risk of a leachable.

Leachables present unique analytical challenges. Since leachables are not related to the drug product, the analytical methods used to detect impurities in the drug product may not be able to detect the leachables.  Even when leachables could be detected by drug product impurities methods, the leachables are often at levels orders of magnitude lower than drug degradation products or related substances, thus below the sensitivity of the method. Thus separate analytical methods are usually needed for the analysis of leachables in the drug product.

Before an analytical method for leachables can be developed, first potential leachables need to be identified. This is done by performing an extraction study on the CCS under exaggerated conditions with the goal of identifying the observed extractables.  Extractables are the compounds that can be extracted from the CCS that might become leachables.  Figure 1

illustrates the ideal relationship between extractables and leachables.

The conditions of the extraction study are selected based upon the drug product and are designed to mimic a “worst case” scenario for the intended drug product. Care must be taken in the selection process so that conditions are aggressive enough to ensure that the extractables include all leachables while not being too aggressive thus generating an impractically large number of extractables that are not leachables. The extraction study should not lead to a complete deformulation of the material.

Study Design

Sample Selection

All components of the CCS that directly contact the drug product either during storage or during the administration of the drug product are considered to be primary components of the CCS.

All components of the CCS that do not contact the drug but do have the potential to interact with the primary CCS are considered to be secondary components of the CCS.  A secondary CCS component will either contract the primary CCS or contact another secondary CCS component that does directly contact the primary CCS.  Table 3 shows some common examples of primary and secondary components of CCSs.

All primary CCS components should be included in the extraction.  If in the final CCS the

components are to be pre-treated in any way (e.g. sterilized) before being filled with the drug product, the samples to be used in the extraction study should be pretreated in a similar manner to ensure that the extraction profile correctly models the CCS exposed to the drug product.

Selection of secondary CCS components for inclusion in the extraction study is based upon a risk assessment. In this risk assessment the likelihood of the secondary CCS component giving rise to leachables and the likelihood of these leachables being able to contact and penetrate the primary CCSC are considered. One secondary CCS component that will usually need to be included in the extraction study is a printed label if it is to directly contact a part of the primary CCS.

Extraction Studies

The first step toward evaluating leachables is to perform extractions studies. There are two types of extraction studies; Controlled Extractions and Simulated use extraction. These two extractions can be done in series or in parallel.  In some cases, just one of the extraction studies may be sufficient.

A Controlled Extraction study (also called materials characterization study) involves extracting the CCS in 2-3 solvents of varying polarities. A CE study using 3 solvents is required for drug products at the highest risk based upon the route of administration in Table 2.  A CE study using 2 solvent is recommended for drug products in the other two risk categories for route of administration if the risk of packaging component-dosage form interaction is high in Table 2.

The solvents are selected based upon the drug product with at least one of the solvents representing a “worst case scenario”. The extraction conditions used are aggressive, typically reflux or oven incubation. The combination of the “worst case scenario” solvent with the aggressive extraction conditions is intended to yield a high number of extractables. The end result of this approach is that all potential leachables (except those that react or have a unique affinity with the drug product) will be identified.  Table 4 shows example extraction solvents for an aqueous drug product.

A simulated use extraction study (also called a simulation study) involves extracting the CCS in solvents that closely mimic the drug product. The extraction conditions are usually static storage of the CCS in the solvent at a temperature above the intended storage condition of the packaged final drug product. The end result of this approach is that the observed extractables are likely to be leachables.

A simulated use extraction is designed to be less aggressive than a controlled extraction study, thus less extractables are expected to be identified in a simulated use extraction compared to a controlled extraction. The simulated use study is more likely to identify only the extractables that will become leachables compared to the controlled extraction study which will potentially identify many extractables that will not become leachables. However, a simulated use study is more likely to “miss” a potential leachable than a controlled extraction study. Both studies reveal useful information on the potential leachables from a given material but the project team must be aware of the strengths and weaknesses of each study.

Regardless of type of extraction study performed, once completed, the sample extracts are analyzed by at least GC-MS, LC-MS and ICP-MS.  The goal of these analysis is to identify as many extractables as possible and to semi-quantitatively determine the level of each extractable.  Based on the material, additional analysis may be required for specific leachables known to be highly toxic.  Since the methods are designed to detect unknowns, these methods cannot be validated. Results from these analyses are reported as the amount of the extractable (usually in µg) per weight (usually in g) or surface area (usually in cm2) of the CCS component.

Analytical Evaluation Threshold

At the completion of the extraction studies, a list of extractables is generated.  The challenge at this point is to select which extractables present a toxicological risk and thus should be monitored as leachable.

To evaluate the toxicity of each observed extractable, the safety concern threshold (SCT) is used. The SCT is the absolute highest acceptable exposure of a patient to a leachable in drug product and is usually expressed in terms of µg of leachable per day.   If an SCT is not known, the recommended SCT by the PQRI should be used.  The PQRI selected the recommended SCT as representing a threshold below which leachables would have negligible safety concerns from carcinogenic and non-carcinogenic toxic effects.  For orally inhaled and nasal drug products (OINDP), the PQRI recommended SCT is 0.15 µg of each individual leachable per day.   For parenteral and ophthalmic drug products (PODP), the PQRI recommended SCT

is 1.5 µg of each individual leachable per day.

To apply the SCT to a given drug product, an analytical evaluation threshold (AET) is calculated based on the SCT of an individual leachable, the number of doses of the drug product administered per day, the number of doses contained in the container closure system (CCS), and the weight of the CCS (can also use volume of drug product in the CCS.

The AET will have units of µg/g unless other units were used in the calculation. Surface area of the CCS or volume of drug product in CCS instead of the weight may be applicable in some situations. The uncertainty factor is an adjustment for the confidence in the identification and quantitation of the extractables needed for OINDP.  For all other types of drug products, the uncertainty factor is not needed.

Selecting Leachables

All extractables above the AET should either be selected to be monitored as a potential leachable or submitted for a toxicological assement to determine a compound specific SCT.

An example set of extractable results are shown Table 5 for a fictional CCS with an AET of 10.0 ppm for each extractable. The results listed in italics are well below the AET and would not be selected as target leachables. The results that have been bolded are significantly above the AET and would definitely need to be selected as target leachables. The results that are underlined represent results that would require additional consideration since the results are close to the AET. The SCT and the uncertainty factor should be reevaluated before selecting or dismissing these extractables as leachables. A conservative selection of including an extractable that is just below the AET as a target leachable is an acceptable and common practice.

Development and Validation of Analytical Methods for Leachables

The goal of the analytical methods is to have sufficient sensitivity so that the LOQ is at or below the AET. Extensive sample preparation may be necessary to ensure sufficient sensitivity.

The analytical methods are then validated with the goal to meet the ICH acceptance criteria. However, since the challenge of these methods is to be able to detect very low levels of leachables in often complex drug product matrices, some allowances may need to be made in other aspects of method performance to allow sufficient sensitivity. These allowances may be seen in higher acceptance criteria than in drug product impurity methods.

Analysis of Leachables in Drug Products

The analytical methods are then used to analyze drug product stored in the CCS under the intended storage conditions. Ideally this testing can be done as part of the stability study but it can also be done in a separate migration study. Results from the analysis are reported as concentration of the leachable in the drug product.  If a leachable is observed above the AET, additional experimentation may be necessary to confirm the identification of the leachable. Additional toxicological evaluation may also be needed to assess the risk of the leachable.

Control and Placebo Samples

Inclusion of control samples can greatly simplify the analysis of the leachables results. One recommended control is to store the drug product in a different CCS under the same storage conditions for the same length of time.  Ideally the different CCS would be expected to yield significantly less and different leachables (e.g. a glass vial with a Teflon coated lid).  This control is used to distinguish degradation products of the matrix from leachables.

Including placebo samples stored in the CCS in the leachable study is strongly recommended. These samples can help to confirm the presence of leachables observed in the active and leachables might be observed in the placebo that might have been missed in the active. One cannot assume, however, that if a peak is observed in the active that is not in the placebo that the peak is not a leachable. The active drug may facilitate the migration of the leachable or the active drug could react with the leachable.

Impact of Leachables and in the Drug Product Above the AET

If a leachable is observed in the drug product above the AET calculated for that specific leachable, the project team must take actions to prevent patients from being exposed to this level of the leachable. The options are the shelf like must be shortened to a time before the leachable exceeds the AET or a different CCS must be selected. If a different CCS is selected, the entire extractables and leachables testing must be repeated on the new CCS.

If no leachables are observed in the drug product above the AET the project team can set the shelf life and storage conditions of the drug product in the CCS based solely on drug product stability. Thankfully this is commonly the case.

Conclusion

Leachables present a unique challenge in assuring drug product safety and efficacy. The    experimental approach discussed in this chapter represents a rational experimental approach to evaluate this risk. When a project team designs experiments based on this approach, the more information the team gathers on the material composing the CCS and on the drug product, the more effective and efficient the experimental strategy will be.

READ
September 5, 2018

Cleaning Validation for Medical Devices: Establishing Cleaning Limits

After the potential hazardous contaminants from the manufacturing process have been identified during the review of the process flow, the acceptable levels for the contaminants is determined.   For contaminants with complete toxicological data available, follow ISO 10993-17 to set the cleaning limits for the contaminants.    When limited toxicological data is available, at a minimum the LD50 values can be used to calculate the cleaning limits.   For most contaminants the LD50 values are readily obtained from the MSDS.

The established cleaning levels are then used in the selection and evaluation of the cleaning test methods.

READ
August 16, 2018

Preparing to Adopt a Quality Agreement at a CRO: Check That Your Quality House is in Order

There should already be in place a strong quality program at the CRO and the quality agreement must make a good fit to this program.  Those concerns (terms) that are rather common to Quality Agreements should already be addressed in the existing quality program at the CRO or else consideration must be given to incorporating them into the existing program.  This approach is distinctly different and better than leaving them only in the quality agreement.  To leave important concerns only in the quality agreement is tantamount to instructing the lab to only consider doing such quality work when working on this particular client’s project.

READ
August 13, 2018

Possible Outcomes of a Leachables Study

At the end of the stability study, results from the leachables analyses will be reported.  There are two basic outcomes to a leachables study:

  1. Good news: At the end of the stability study, all leachables were below the AET.
  • No action needed
  • Container closure system has no impact on shelf life

  1. Bad News: One (or more) leachable exceeded the AET at or before the intended shelf life.
  • Assess toxicity of the leachable to determine if SCT was appropriate. If SCT increases after assessment, AET can increase.
  • If after confirmation and toxicity assessment the leachable is still above the AET, shelf life must decrease to time when leachable was below AET.
  • In worst case scenario, a new container closure system may be needed.